6.1.2 A2 - ALARMS Limits : respiratory parameters

Touching this icon, a window showing the Alarm limits is displayed: the User can possibly set the Min
and Max values of each single alarm.
Morpheus_ND
WARNING !! Risk for Patient / User injury
The anaesthesia unit used in the same health environments can have different
preset configurations of alarm limits.
Verify that the preset alarm limits are appropriate for the new patient and adjust
the alarm limits on values suitable to the new condition of use.
The alarms setting to the range limit could make the alarm not properly
working.
Select: ALARMS icon
Nota. The alarm limits can be modified
also while the equipment works.
The Alarm Limits screen appears.
Select
YES: to quit Alarm page; the alarm set
will NOT be saved.
NO: it remains in Alarm page.
Select
YES: to quit Alarm page; the alarm set
will be saved.
NO: it remains in Alarm page.