5.8.6 VC-VAC BABY

Volume targeted controlled ventilation synchronised with
the patient if the inspiratory trigger for paediatric and
neonatal patients is enabled.
The system displays all PRP related to the operative
mode set.
VC/VAC BABY
breaths if the inspiratory trigger (Tr. I) is enabled.
The Operative Mode VC/VAC BABY is active only with patients: PAEDIATRIC and NEONATAL.
In this type of ventilation the work of breathing is fully assumed by the unit, and therefore it is used
when the patient is unable to breath autonomously, or in order to assure an efficient pre-set current
volume and therefore the mechanical ventilation must fully replace the spontaneous breathing.
5-40
is a volume-targeted controlled ventilation (Vte), synchronised with the patient's
The VC/VAC BABY operative mode also includes two additional parameters
identifying the maximum and the minimum pressure limit (PMax. - Pmin) that can be
reached during ventilation.
User manual, DU33ND101