O2 Sensor Calibration; O 2 Sensor Calibration - Siare Falco 202 Evo User Manual

Lung ventilator
Table of Contents

Advertisement

3.4.3 O
sensor calibration
2
WARNING !! Risk of M.D. failure and/or injuries for the patient
Select O
2
calibration.
The system
automatically will enable
the O
sensor
2
calibration.
To determine the correct
operation of the O
sensor, the software
reads the electrical value
(mV) generated by the
cell when è in presence
of a flow with a 21% O
concentration (Air) .
The O
sensor
2
calibration in progress.
At the end of the O
sensor is new and in perfect conditions): Test Completed ( XXmV ) which
show the value in Volts measured by the sensor with 21% Oxygen.
3-24
This procedure should be carried out to check the proper operation of the
oxygen sensor.
Perform this procedure monthly.
Sensor
2
2
sensor calibration a message is showed (if the O
2
2
User manual, DU3104101