2.2.1 Gas supply connection

WARNING !! Risk of equipment failure
WARNING !! Fire danger
CAUTION
Morpheus_ND
Connections for O
Connections for O
Connections for Air entry
Connections for N
In order that the anaesthesia unit operates as specified, the inlet pressures of
medical gases must be between 280 kPa and 600 kPa (2.8 - 6 bar): before use
ensure that this requirement is met.
The medical gas source shouldn't contain water: if you suspect the presence of
water, connect a water trap to avoid damages on the unit and its components.
Do not connect to the connector for O
guaranteed to operate with pure oxygen.
Do not execute connections to the medical gas distribution system or to
cylinders before having consulted the relevant chapter (see on chapter 4.3).
With the anaesthesia unit operating by TURBINE, it is anyhow necessary the
presence of an oxygen supply, useful for the operating of the internal pneumatic
system.
The gas used must be of medical type, therefore oil free and filtered.