6.2

Alarms setting

6.2.1 Setting of ALARMS limits values

The User can modify and display the Alarm Limits values in two different ways; through the ALARMS
selection icon or selecting a Respiratory Parameter icon.
Morpheus_ND
WARNING !! Risk for Patient injury
Using in the same area more than one medical device having different alarm limits
setting, the User could have a potential false misinterpretation.
Before using the anaesthesia unit it is suggested to adjust the parameters
necessary for the correct operation of the anaesthesia unit.
During operation it is possible to adapt the ALARM limits values setting in
function of the patient clinical situation.
Select: ALARMS icon
The Alarm Limits screen appears.
Select
YES: to quit Alarm page; the alarm set
will NOT be saved.
NO: it remains in Alarm page.
Select
YES: to quit Alarm page; the alarm set
will be saved.
NO: it remains in Alarm page.
6-11