3.4.3 List of available respiratory parameter

WARNING !! Risk for Patient injury
Depending on the chosen ventilation mode, the same PRP can be a dependent
variable (that varies depending on other parameters modification) or an independent
variable (a value that if modified, affects the values of other parameters).
The PRP must be set by the User in Stand-by mode before activating the necessary
operative mode.
The PRP can also be adjusted while the anaesthesia unit runs, adapting them to the
patient's clinical situation.
Selecting the PRP parameters icon the parameters relevant to
the set Operative Mode are displayed.
The parameters marked with BK are referred to the BACK-UP
operative mode.
RR BK (bpm) : Back-up Respiratory Rate, used when an
apnoea condition arises to enable a controlled ventilation mode.
3-26
I:E
Ratio between inspiration and expiration phases.
Pause (%)
Inspiratory pause time. The "inspiratory pause time" is displayed on the
monitor in % (% of the inspiratory time). It is also used to calculate the lung
mechanics parameters (resistance and static compliance).
The value of Respiratory Rate has
been modified.
Confirm
Select
to close
modification bar
User manual, DU33ND101