Tables Of Alarm Characteristics; Alarms Configurable By User - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

6.3

Tables of alarm characteristics

6.3.1 Alarms configurable by User

Alarm type
LOW PRESSURE
OF AIRWAYS
HIGH PRESSURE
OF AIRWAYS
LOW
RESPIRATORY
RATE
HIGH
RESPIRATORY
RATE
LOW EXP. VOLUME
HIGH EXP. VOLUME
LOW VOLUME
MINUTE
HIGH VOLUME
MINUTE
LOW PEEP
HIGH PEEP
LOW FiO
2
HIGH FiO
2
LOW EtCO
2
HIGH EtCO
2
Apnea
POWER SUPPLY
FAULT
Morpheus_ND
Priority
Activation
delay (sec.)
HIGH
15
HIGH
0
patient
3
HIGH
breaths
patient
3
HIGH
breaths
HIGH
15
HIGH
0
HIGH
60
HIGH
60
HIGH
0
HIGH
0
HIGH
30
HIGH
30
HIGH
30
HIGH
30
HIGH
5 - 60
HIGH
15
Suspendable
Suspension
delay (sec.)
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
Inhibition
15
NO
-
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
120
YES
6-19