Periodic Maintenance; Maintenance Operations Table - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

7.4

Periodic maintenance

Inspections and periodic maintenance are ensured by taking out a maintenance contract with
SIARE or an authorised dealer.
Contact SIARE for information regarding authorised Service Centres in your area.
When you require service, please indicate the serial number of the unit and the problem to
SIARE or to your authorised technicians.
SIARE assumes responsibility for all provisions foreseen by the law, if the equipment is used
and maintained as per the instructions in this manual and the technical manual.
The Technical Assistance Report, signed by the authorised SIARE technician, is proof of the
completion of the scheduled maintenance.

7.4.1 Maintenance operations table

The following table summarizes the preventive maintenance frequency and procedures to be carried
out on the anaesthesia unit.
Morpheus_ND
Disposable bacteria filter.
Replace the bacterial filter as per manufacturer's advice (24 - 72
hours).
If used with humidifier, change the filter more frequently (as and
when the resistance increases or patient's "work of breathing"
increases).
Immediately replace the filter if it is stained with blood,
secretions or other human fluid
Check the expiry date before using on patient.
Change the filter early if nebulisation is done frequently.
The anaesthesia unit does not require particular maintenance and preventive
operations other than those indicated in this manual or in order to respect
standards applied in the specific country where the unit is sold.
WARNING !!
Risk of injury for the patient
Always refer to the instructions contained in the previous section: cleaning,
disinfection and sterilization of the components.
7-9