Preliminary Checks; Introductions - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

4.9

Preliminary checks

4.9.1 Introductions

The preliminary checks are divided in 4 sections
SUPPLEMENTARY TEST
• O
Sensor Calibration
2
• Leak Test
FLOWMETER
VENTILATION MODULE
• Physiological respiratory parameters
• Parameters monitoring
• Anaesthesia unit alarms
OPTIONAL
WARNING !! Risk of unit failure and/or injuries for the patient
Running or cancelling the preliminary checks might result in a malfunction during
ventilatory operation: pay utmost attention.
Always carry out all preliminary checks, unless there is an emergency situation.
The User should carry out the preliminary checks once the emergency condition
stops, and at least once a week.
CAUTION
The anaesthesia unit must be ready for use in order for you to start with the
preliminary checks. Proceed with the following operations / verifications.
4-28
Connect, electric power supply, medical gas and patient circuit.
Insert and connect, oxygen sensor.
Insert, CO
canister and Manual ventilation KIT
2
Connect a patient simulator to the patient circuit terminal.
Anaesthesia unit ON : Stand-by mode.