Meaning Of Lung Status Icon - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

3.6.5 Meaning of lung status icon

The lung status icon simulates the patient's lungs, graphically displaying the
respiratory cycle by alternatively switching the lungs color.
In fact in case of patient's spontaneous activity ( Trigger ), the lung status icon turns
to yellow and if the " LOW PRESSURE" alarm value set has not been exceeded the
icon turns red.
The lungs turn green during inspiration only if the PAW exceeds the " LOW
PRESSURE" alarm value set.
WARNING!! Risk for Patient injury.
If the airways pressure does not reach the value of " LOW PRESSURE " alarm
parameter, the lungs icon turns red and after about 15 seconds, the system activates
the Low Pressure alarm.
Morpheus_ND
During inspiration the lungs icon turns green.
During expiration the lungs icon turns light grey.
At trigger activation the lungs icon is yellow.
If the airways pressure does not reach the value of " LOW
PRESSURE " alarm parameter. the lungs icon turns red.
3-45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents