Leak Test Not Overcome - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

4.9.4 Leak Test not overcome

WARNING !! Risk of anaesthesia unit failure
The Leak Test is not overcome when the pressure inside the pneumatic circuit does
not reach al least 27 cmH
CAUTION - Leak Test ERROR
Verify that :
Repeat the Leak test.
Leak Test ERROR
Please see chapter 6 : Alarms and Troubleshooting.
If the problem persists, contact the Siare Technical Centre or a Technical Centre
authorized by Siare.
4-32
O.
2
the anaesthesia unit is in STAND-BY
the patient circuit is correctly connected to the flow sensor and to the INSP
connector (on breathing system)
Y of patient circuit is correctly closed / plugged
the anaesthesia unit delivers a flow (check the breathing system).
Leak Test procedure
successfully completed: the measured
pressure value is 14 cmH2O.
User manual, DU33ND101