7.3.2 Disposable bacteria filter
Disposable
bacteria filter
7-8
The following are generic agents recommended for cold
sterilization of EXP flow sensor.
Cidex Plus
Cidex OPA
Denatured Ethyl Alcohol
70% Iso-Propyl Alcohol
5.25% - 6.15% Sodium Hypochlorite solution
Sporox II (Hydrogen Peroxide)
Protex wipes (Octyl decyl dimethyl ammonium chloride, Dioctyl
dim ethyl ammonium chloride, Didecyl dimethyl ammonium
chloride, Alkyl dimethyl benzyl ammonium chloride)
Hexanios G+R
Liquid alkaline cleaner
CAUTION
It is important to use a disposable bacteria filter at the expiratory port
of anaesthesia unit between the patient circuit's expiratory limb and
the expiratory flow sensor.
Do not clean or re-use if the filters are the
disposable type.
WARNING!! Risk of device failure
Disposable bacteria filter.
Do not attempt to sterilize and reuse a disposable bacterial
/HEPA filter. It is meant for single use.
Do not use filters that aren't in sealed pack.
Do not try to dry the filter and reuse in the same patient in case
of increased resistance.
Do not use it longer than the time-period recommended by
manufacturer.
Do not use filters that had exceeded the expiry date.
Components that cannot be
destroyed should be sterilized
and disinfected according to local
standards.
User manual, DU33ND101