6.1.5 A5 - Acoustic alarm silencing

During the normal operating phase of anaesthesia unit, it is
possible to silence the active acoustic alarm.
By pressing the Alarm Reset soft key.
By selecting the bell evidencing the active state of the alarm.
6-10
WARNING !! Risk for Patient / User injury
The User should never stop checking the patient conditions during the alarm
silencing.
The acoustic alarm silencing is active and will stop the acoustic alarm for a
defined time.
During the alarm silencing the alarm text is showed on the panel.
A new alarm silencing will cancel the alarm text only if the alarm conditions is
disappeared.
If during the alarm silencing a new high priority alarm occur, the alarm silencing
is cancelled and the acoustic signal and the visual texts are activated again.
CAUTION
It's possible to change the Alarm limit settings even when the alarm is activated.
After changing an alarm limit, the relevant sign is lighted and the status icon will
flash for a defined time.
User manual, DU33TS101