4.3.7 Gas analyzer connection
•
Extract the gas analyzer and the cable +
interface module, from package.
•
Switch OFF the anaethesia unit power supply
- OFF
•
Connect the interface cable between the RS-
232 connector (positioned on the side of
anaethesia unit) and the interface module.
•
Connect the GAS ANALYZER to the interface
module.
Mainstream GAS ANALYZER
•
Apply the gas sensor to the patient circuit.
Sidestream GAS ANALYZER
•
Connect the sampling line to the GAS
ANALYZER.
•
Connect the sampling line to the patient
circuit.
Verify the functioning of the Mainstream GAS ANALYZER
•
•
Verify the functioning of the Sidestream GAS ANALYZER
•
•
Morpheus_ND
For further information on operating logic and on gas sensor malfunctioning, make
reference to the gas analyzer manual supplied with the device.
Switch ON the electric power supply.
A green LED ( on the gas analyzer ) indicates that the IRMA analyzer is ready
for use.
Switch ON the electric power supply.
A green LED ( on the gas analyzer - sampling line ) indicates that the ISA
analyzer is ready for use.