4.5.3 Battery power supply

The operation time of the anaesthesia unit guaranteed by the battery can vary in the following
cases: old battery or not fully efficient, unusual ventilatory parameters, presence of electronic
flowmeter box.
Replace the battery each two years ( please refer to chapter 7 Maintenance ).
WARNING !! Risk of personal-physical injuries
To silence the acoustic alarm,
push the key " ALARM RESET ".
When the led on the front panel is
on it indicates that the
anaesthesia unit is correctly
supplied.
Battery charger
The anaesthesia unit battery can be recharged by leaving the anaesthesia unit
connected to the main power supply (using power cable supplied with the unit) it is
not necessary that the unit is on. For battery recharging follow what specified at
chapter 4.2.3.
WARNING !! Risk of failure
In case of power failure, the optional devices connected to the supplementary
electric socket are not not supplied by the battery. Pay much warning to all power
supply indicators of connected devices.
Morpheus_ND
The battery, inside the anaesthesia unit, must always be installed.
If the battery is not fitted, the anaesthesia unit is not protected against
fluctuations or failures of the power supply.
The anaesthesia unit must not be used without a mains power supply.
The use of the battery must be limited to short periods and is not foreseen as an
alternative to the mains power supply.
Do not open the mains power supply module to replace the battery or to carry
out maintenance operations on the battery charger.
4-17