2.7
Breathing system use
CAUTION
•
•
2.7.1
CO
soda lime absorber canister
2
The CO
absorber canister is positioned in the upper side of valves group; it can be easily
2
disconnected by apposite lock lever, this functionality makes it replaceable also during surgical
interventions.
•
When the CO
rebreathing modality.
•
When the canister is removed, the system is automatically configured in non-rebreathing
modality (real open circuit).
•
It is possible to insert and remove the canister during interventions. The canister is available in
reusable or pre-loaded disposable versions.
About assembling and disassembling of absorber canister see on next chapter.
2.7.2
Assembling and disassembling of absorber canister (e.g. disposable model)
5-12
When the CO
absorber canister is inserted, the system is automatically
2
configured in rebreathing modality.
When the CO
absorber canister is removed, the system is automatically
2
configured in non-rebreathing modality.
absorber canister is inserted, the system is automatically configured in
2
•
To unpack carefully the CO
•
Shake the absorber canister (disposable model) in order to separate
the soda lime granules.
•
Remove the seal from absorber canister (disposable model).
•
Insert the absorber canister in the apposite groove (opposite side of
the "PUSH" lever).
absorber canister.
2
User manual, DU33ND101