4.5

Connection to power supply

The electrical connections are a very important part in the installation of the anaesthesia unit.
Incorrect connections or connections to unsuitable electrical systems can compromise the safety of
the patient and the User. The mains power supply must comply with the CEI 64-8/7 regulations
concerning type A premises for medical use.
The power supplies foreseen on MORPHEUS_ND anaesthesia unit are of three types:
by main power supply ( 100 - 240Vac / 50 - 60Hz )
by low voltage power supply ( 12Vdc / 7A )
by internal battery ( NiMh 4,5 Ah 12Vcc : max. autonomy 2 hours).

4.5.1 Main power supply

The main electric power supply must correspond to that indicated on the identification label (electric
power supply, frequency and power consumption) located on the back side of anaesthesia unit: 100 -
240Vac / 50 - 60Hz / 120 VA.
WARNING !! Patient / User injury hazard
SIARE specialized personnel or qualified technical personnel formerly authorized by
SIARE, shall always verify the following during the installation.
Connect the power cable outlet to the plug on the
anaesthesia unit.
Insert the plug of the power cable in the main
power supply outlet.
Position the main switch (located in the back side
of the anaesthesia unit) in position "I".
Verify on ventilator module front side if the green
led is lit (it indicates the presence of main power
supply).
Morpheus_ND
The presence of the ground cable in the electric plug used to connect the
anaesthesia unit
The electric plug used should be equipped with a lock in order to prevent
erroneous placements in wrong plugs without ground.
4-15