LEAK TEST

Based on known and calculated data (flow, pressure and time), the software
calculates the parameters displayed at the end of test. The value of the compliance
obtained during the Leak Test is used for the "compensation of dead space" function
( e.g. 0,3 x cmH
Range of values accepted by the Leak Test
Patient Circuit Compliance
Leak (ml/min)
Morpheus_ND
O measured ).
2
Leak Test
The
procedure is in
progress.
Leak Test procedure
successfully completed: the measured
value is 9 ml/min.
Leak Test
If the procedure of
been correctly performed the
Test Completed
message "
displayed.
In case the test has not passed, one
of the following messages is
displayed.
" Test Aborted "
" Test Failed "
Min.
0,2
0
has been
has
" is
Max.
4
100
4-31