O2 Sensor Calibration - Siare Falco 202 User Manual

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

3.6.2 O
Sensor calibration
2
TESTS ON DEMAND. O
screen.
Falco 202
CAUTION - TESTS ON DEMAND
The TESTS ON DEMAND screen shows various tests.
Turbine Characterization
Expiratory Flow Sensors Calibration
O
Sensor Calibration
2
On this TEST ON DEMAND phase you should consider the O
calibration while for the other tests please refer to chapter 3.8.
WARNING !! Ventilator malfunctions risk
This procedure should be carried out to check the proper operation of
the O
sensor.
2
To avoid any patient injury risks, carry out this procedure monthly.
sensor calibration
2
Press the knob selecting,
TESTS ON DEMAND.
The TESTS ON DEMAND
screen appears.
Sensor
2
3-21