8.3.3

ANNEX C: Table 3

Guidance and manufacturer's declaration – electromagnetic immunity
The Morpheus is intended for use in the electromagnetic environment specified below. The customer or
the user of the Morpheus should assure that it is used in such an environment.
IEC 60601 TEST
IMMUNITY test
3 Vrms
150 kHz to 80 MHz
outside ISM bands
Conducted RF
IEC 61000-4-6
10 Vrms
150 kHz to 80 MHz
in ISM bands
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Morpheus_ND
Compliance
LEVEL
level effective
VRMS
VRMS
V/m
Electromagnetic
environment – guidance
Compliance
level
Recommended separation
3 Veff
SEE Annex E
10 Veff
SEE Annex E
80 ÷ 800 MHz
SEE Annex E
10 V/m
800 ÷ 2500 MHz
SEE Annex E
distances
8-19