2.7.3 Dismount of the valves group

WARNING !! Risk of malfunctioning
All the interventions of maintenance on valves group, must be affected exclusively
by personnel highly qualified and specifically trained and formally authorized by
SIARE. Pay much ATTENTION during this operation.
Morpheus_ND
Press the apposite lock "PUSH" lever (unlock).
Press down and release the apposite lock lever (unlock).
Push (lever side) down the absorber canister.
Release the lock "PUSH" lever (unlock).
To take off the absorber jar, press the "PUSH" lever.
Pull up (lever side) the CO2 absorber canister.
Extract from the apposite groove (opposite to "PUSH" lever side).
Dismount the eventual circuits or bags connected to the valves group
connectors.
Pull up the valves group cover.
Pull up the handle.
Extract the valves group, pulling it up vertically.
The pneumatic connections are automatic with double tightness
gaskets to prevent accidental leakages.
Lift and remove completely the valves group and position it on a flat
surface.
Continue with the Maintenance (see on chapter 7).