2.1.2 Other commands

Back-lit
flowmeter
Fresch gas
control knob
APL valve
( manual
ventilation )
Manometers
CAUTION - In case of emergency.
In case of emergency a knob delivering a monitored oxygen flow (
lit mechanical flowmeter (safely calibrated) and indicating the oxygen delivered is
foreseen (
Morpheus_ND
The flowmeter and the knob are to be used mainly in an emergency. In fact, in
the event of an emergency, with the anesthesia unit switched off
(malfunctioning), the User can deliver a flow of fresh gas (oxygen) adjustable
through the AUX outlet (located on the back of the DM) for manual ventilation of
a patient.
APL valve
must be used
manual ventilation
The pressure value increases by turning clockwise the valve knob and it
decreases by turning it counter-clockwise
The adjustment range is around from 0 to 50 cmH
The three manometers tell the User the medical gases feeding pressure.
FRESH GAS
)
for adjustment of airways maximum pressure during
O
2
O
) from a back-
2
2-3