Section 8

Instrument

Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Perform the following steps prior to shipping/relocating the
instrument or prior to long-term shutdown (more than two (2)
months).
1. Power OFF the Abbott VP 2000 Processor. Unplug the
Abbott VP 2000 Processor from the power source (UPS or
outlet).
2. Power OFF and disconnect the PC from the Abbott VP
2000 Processor.
3. Drain and remove the Fluid-Flow Lines (Drain Line and
Water Line).
4. Remove the Slide Basket from the Robotic Arm.
5. Remove the Hoods.
6. Remove the Vessel Carrier.
7. Move the Robotic Arm all the way to the left side of the
unit (the side next to the power switch).
8. Lock the Robotic Arm against the left vertical support
using the internal shipping strap supplied with the unit.
WARNING: Do not ship the unit without locking the
Robotic Arm securely in place. Otherwise, the internal
drive mechanism will be damaged.