Removal - Abbott VP 2000 Service Manual

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Section 7

Removal

Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
1. Remove the Hoods by gently squeezing inward on the two
(2) metal arms that connect the hoods to the unit.
2. Remove the Top Cover by unscrewing six (6) 6-32 x 1/4
Phillips-Head screws, three (3) on each side of the unit.
3. Push the Tension Arm toward the X-Motor to relax the
belt tension; remove the drive belt from the motor pulley.
4. Remove the X-Sensor Board by unfastening sixteen (16)
6-32 x 1.4 Phillips-Head screws. (The X/Y Table Assembly
can be shifted left and right to reach the screws.) Slide the
X/Y Table Assembly to the right and carefully pull the left
end of the X-Sensor horizontally and away from the frame
in order to disengage the J4 connector (on the not-visible
side of the X-Sensor Board). Ensure that the J4 connector
is completely disengaged; then remove the X-Sensor Board
by sliding it to the end of the unit containing the power
switch.
X-Sensor Board Assembly
7-23