Section 5
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
14. Push "TEST" - "RUN" - "1" - "ENTER". A built-in test will
start. When the arm moves past Bath 2, place the power
switch in the OFF position. The power interrupt alarm
should continue to sound. The arm will return to Bath 1.
Push "CLEAR" and the power interrupt alarm should go
OFF.
Check off the following block on the checklist:
POWER INTERRUPT ALARM
15. Place the UUT power switch in the ON position. When the
control panel displays "READY" press "TEST" - "ENTER"
keys. Next, press the following keys and observe the
number symbol that appears in the Bath display. Verify
each number.
Table 5.4:
Bath Display Number Symbols
Key
PROG
REVIEW
PAUSE
RUN
AGITATE
PRINT
ABORT
UP ARROW
LEFT ARROW
RIGHT ARROW
DOWN ARROW
1
4
7
CLEAR
2
Number
13
20
16
14
19
15
21
22
25
24
23
1
4
7
12
2
Electrical Test
5-41