Maintenance and Diagnostic Procedures
Manually Controlling the Basin Heaters
5-34
There may be times when the basin that is not being used in
the current protocol needs to be pre-heated. Use the following
procedure to manually turn a basin heater(s) ON. The
procedure can be performed at any time during a run or when
the system is idle.
1. Press the <Run Screen> button on the left-side of the
window to display the "Run Screen" window.
Figure 5.22: Run Screen Window
Section 5
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008