Water Bath Data; Drying Station; Performance Data - Abbott VP 2000 Service Manual

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Technical Data

Water Bath Data

Input Line Pressure
Water Type
Flow Thru Rate
Recirculation Rate
Water Bath Capacity
Water In Line
Drain Line

Drying Station

Temperature

Performance Data

Slide Capacity
Unheated Reagent Vessels
Heated Reagent Vessels
Water Bath
Drying Station
Reagent Basin Protocol Volume to Cover Slide
Reagent Basin Protocol Volume to Cover Sample
Heated Reagent Temperature Adjustment
3-28
Description
10 to 100 psi
Deionized
Factory set via regulator
approximately 1 liter/min
Approximately 3.5 liters/min
Approximately 1,750 ml
3/8 inch reinforced flexible tubing
to fit standard laboratory water tap.
Adapter required for other taps.
1/2 inch flexible tubing gravity feed.
The entire line must be below the
water level in the unit.
Table 3.6:
Description
Table 3.7:
Description
Table 3.5:
Water Bath Conditions
Specification
Drying Station
Specification
o
Adjustable 25
C to 80
Performance Data
Section 3
o
C
Specification
50
12
3
1
1
600 ml
470 ml
Basin 1: Ambient to 80
Basin 2: Ambient to 80
Basin 3: Ambient to 40
Abbott VP 2000™ Processor Service Manual
30-608308//R1—September 2008
o
C
o
C
o
C