Repair, Replacement of Spare Parts

Removal

Installation

7-64
1. Remove the Hoods by gently squeezing inward on the two
(2) metal arms that connect the Hoods to the unit.
2. Remove the Top cover by unscrewing six (6) 6-32 x 1/4
Phillips-Head screws, three (3) on each side of the unit.
3. Loosen the four (4) 6-32 x 3/4 Phillips-Head screws
fastening the Y-Motor Assembly to the mounting bracket.
Slide the Y-Motor Assembly toward the X/Y Table to relax
the belt tension. Remove the Y-Drive Belt.
4. Loosen the two (2) set screws on the bottom of the Y-Drive
Pulley and remove from the shaft
Reassemble by following the removal procedure in reverse.
CAUTION: Before sliding the Top Cover back into
position, make sure all wiring, cables and connectors
are replaced snugly, within the unit and not hanging
out where they might be sheared off when replacing
the cover.