Maintenance and Diagnostic Procedures
5-46
27. Place the Robotic Arm in the up position at front of the
slide stainer, verify multimeter readings at:
U15
Pin 11 Of5v
U15
Pin 17 Of < .2v
U15
Pin 16 Of5v
U15
Pin 15 Of < .2v
28. Place the Robotic Arm in the down position at the front of
the slide stainer, verify multimeter readings at:
U15
Pin 11 Of < .2v
U15
Pin 17 Of5v
U15
Pin 16 Of5v
U15
Pin 15 Of < .2v
29. Place the Robotic Arm in the up position at the rear of the
slide stainer, verify multimeter readings at:
U15
Pin 11 Of5v
U15
Pin 17 Of < .2v
U15
Pin 16 Of < .2v
U15
Pin 15 Of5v
30. Place the Robotic Arm in the down position at the rear of
the slide stainer, verify multimeter readings at:
U15
Pin 11 Of < .2v
U15
Pin 17 Of5v
U15
Pin 16 Of < .2v
U15
Pin 15 Of5v
.05v
.05v
.05v
.05v
.05v
.05v
.05v
.05v
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Section 5