Abbott VP 2000 Service Manual page 134

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Maintenance and Diagnostic Procedures
5-44
"ENTER"
"3"
"ENTER"
"5"
"ENTER"
"PROG"
19. Next, press the following sequence of keys:
"RUN"
"1"
"ENTER"
"ENTER"
20. Verify that the previously entered program is executed as
shown below and that the arm returns to Bath 1 at the end
of the program.
STAYS AT BATH 1
GOES DOWN IN BATH 1 FOR 5 SECONDS
COMES UP OUT OF BATH 1
GOES TO BATH 2
GOES DOWN IN BATH 2 FOR 5 SECONDS
COMES UP OUT OF BATH 2
GOES THROUGH TWO SHAKE CYCLES
GOES TO BATH 3
GOES DOWN IN BATH 3 FOR 5 SECONDS
"BEEPER" SIGNALS 5 BEEPS FOR END OF PROGRAM
COMES UP OUT OF BATH 3
GOES THROUGH TWO SHAKE CYCLES
RETURES TO BATH 1
Check off the following block on the checklist:
STORED PROGRAM FUNCTIONAL