Fluid Will Not Recirculate: Fault 14.0 - Abbott VP 2000 Service Manual

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Troubleshooting

Fluid Will Not Recirculate: Fault 14.0

6-36
1. Power OFF, top cover, center cover, and AC power barrier
removed.
Check that all plug-in connectors [J1, J2, J3, J7, J8, J9, J10,
J11, J12, J13 and J16] are correctly aligned and fully
seated. A disconnect at J16 will prevent the pump from
operating.
WARNING: Electrical Shock Hazard. Power ON. With
the J16 unplugged, check for 120V AC between Pins 1
and 2 when Robotic Arm in Bath 2 and "Recirculate"
button is activated.
2. Power OFf, top cover removed.
Test by substituting known good units for power supply
board and computer board.
3. Replace pump assembly according to Section 8,
Repair/Replacement of Spare Parts.