Section 6
READJUST
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
FLUID WILL NOT FILL RINSE TANK
FLUID LINE HOOKED UP TO
WATER SUPPLY
WATER SUPPLY ON!
YES
FLOW BUTTON ON
(DISPLAY PANEL)
YES
ROBOTIC ARM IN BATH 2 & DOWN
YES
WATER FLOWING INTO TANK
NO
SOLENOID FUNCTIONING
YES
3
REGULATOR SET PROPERLY
NO
YES
REPLACE FLUID-IN CONTROL
COMPUTER BOARD
TEST BY SUBSTITUTION
POWER SUPPLY BOARD
TEST BY SUBSTITUTION
Figure 6.14: Fluid Will Not Fill Rinse Tank
Maintenance Diagnostic Faults
NO
CONNECT FLUID LINE TO SOURCE
CHECK FLUID LINES FOR KINKS &
BLOCKAGES
NO
ACTIVATE FLOW BUTTON
(RED LIGHT WILL GO ON)
YES
NO
MOVE TO BATH 2 & DOWN
YES
CHECK FOR PROPER
INSTALLATION OF STANDPIPE
1
NO
CHECK CONNECTORS
PASS
FAIL
4
ASSY
5
5
NO
REPLACE CONTROL
MODULE
2
REALIGN & PLUG IN
6-39