Repair, Replacement of Spare Parts

Installation

7-40
Reassemble by following the removal procedure in reverse.
NOTE: When reinstalling the X/Y Table Assembly on its
rails, position the table so that the X-Position Flag (the
protruding metal tab on the X/Y Table that engages the
sensors on the X-Sensor Board) is between sensors to
prevent damage to the flag. make sure the X-Drive Belt
is above the bottom guide wheel and shoulder screw
heads.
NOTE: When reinstalling the two (2) 6-32 x 3/8
Phillips-Head screws to fasten the upper rail to the
Drive Plate, slide the X/Y Table back and forth to ensure
there is no binding.
NOTE: Remove all the twist from the X-Drive Belt
before fastening it in place. Ensure that there are no
twists in the X-Drive Belt when it is installed over the
motor and idler gears.
CAUTION: Before sliding the Top Cover back into
position, make sure all wiring, cables and connectors
are replaced snugly, within the unit and not hanging
out where they might be sheared off when replacing
the cover.
Section 7
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008