Interface Cable Assembly; Removal - Abbott VP 2000 Service Manual

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Section 7

Interface Cable Assembly

Removal

Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Prior to disassembly, move the Robotic Arm so that it is over
the Water Bath and leave it in the up position.
WARNING: Always disconnect the system power at the
plug to the wall socket before beginning disassembly of
the unit.
Figure 7.35: Interface Cable Assembly
1. Remove the Hoods by gently squeezing inward on the two
(2) metal arms that connect the Hoods to the unit.
2. Remove the Top cover by unscrewing six (6) 6-32 x 1/4
Phillips-Head screws, three (3) on each side of the unit.
3. Remove the four (4) 6-32 x 1/4 Phillips-Head screws at the
top of the right-hand vertical side panel just above the fan
guard. These screws secure the side panel to the frame and
to the top rail.
4. Unplug the Interface Cable Assembly from:
Interface Cable Assembly
7-65