Section 7

Installation

Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Figure 7.9: Control Board to Interface Cable Connection
10. Lift the Heater Module Control Board Cover to access the
Sensor Bar Interface Cable.
11. Disconnect the cable from the Control Board.
1. Lift the Control Board Cover, insert the black connector of
the cable through the opening on top of the cover.
2. Carefully connect the black connector onto the board.
Ensure that NO pins are visible around the connector.
3. Ensure the cable does NOT cover the interlock switch
wires. This will cause the cover to not close correctly.
4. Lower the cover into position, ensuring no wires are
damaged in the process.
5. Continue by following the removal procedure in reverse.
Sensor Bar Interface Cable
7-21