Abbott VP 2000 Service Manual page 140

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Maintenance and Diagnostic Procedures
5-50
Check off the following block on the checklist:
FLOW RATE ADJUST
FLOW ACTION FUNCTIONAL
5. Next, press the "RECIR" button and check that water is
re-circulating in the rinse tank and that there are no leaks.
Leave in this condition for 30 seconds.
Check off the following block on the checklist:
RECIRCULATE FUNCTIONAL
6. Turn OFF water supply. Remove the overflow tube and
allow the water to drain into the bucket. Press the "FLOW"
button and wait until no more water enters the rinse tank.
Next, press the "RECIR" button and wait for the same
conditions. Press the "RECIR" button once more.
7. Turn OFF the UUT and disconnect the hoses. Drain the
system of water.
Check off the following block on the checklist:
DRAIN ACTION FUNCTIONAL
Section 5
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008

Advertisement

Table of Contents
loading

Table of Contents