Section 7

Drain Tube O-Ring

Removal

Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Prior to disassembly, move the Robotic Arm to the far end of
the unit away from the Dryer Station and lift the hood up and
out of the way.
WARNING: Always disconnect the system power at the
plug to the wall socket before beginning disassembly of
the unit.
Figure 7.38: Drain O-Ring
1. Loosen the single thumbscrew on the front face of the unit
between the Drying Station and Bath Station; then,
remove the stainless steel cover plate over the two (2) wells
by lifting it upward and out.
2. Grasp the Drain Tube and gently twist and pull it upward
to remove.
3. Remove the O-Ring at the base of the unit.
Drain Tube O-Ring
7-71