Safety

Safety of the Overall System

Instrument-related Hazards and Safety Measures

2-8
In addition to safety instructions included in this manual,
safety instructions of the overall system must also be observed
and followed.
Follow the safety instructions of all instruments used in
conjunction with the Abbott VP 2000 Processor.
Observe the following safety notices.
CAUTION: Electrostatic Discharge Hazard. Observe
precautions for handling electrostatic-discharge-
sensitive devices.
Wear a wrist ground strap.
Discharge static electricity from your body before
touching the instrument.
WARNING: Pinch Point. The instrument requires
accurate positioning of system modules and slides.
When operating the instrument, it could be necessary
to run tests with the instrument's safety panels open.
Keep body parts away from moving parts of the
instrument (e.g., Robotic Arm) during setup and test
procedures.
Stand back from the worktable surface during
these procedures.
Fast moving devices could cause injuries.
WARNING: Electrical Shock Hazard. Possible danger
from electrical shock. Switch off the main power and
disconnect the main power connection of the
instrument before replacing fuses, printed circuit
boards, major sub-assemblies, etc. Replace only fuses
that are externally accessible and labeled. Only use
replacement fuses of the specified type and electrical
rating.
Section 2
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008