Assay Characteristics; Instrument Troubleshooting - Abbott AEROSET Troubleshooting Manual

Table of Contents

Advertisement

Flowcharts
Abbott Assay—Basic Assay Troubleshooting, Continued

Assay Characteristics

Instrument Troubleshooting

248
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
If multiple assays are affected, look for similar characteristics
such as:
Same line
Assays using R1 only vs. assays using R1 and R2
Similar wavelengths
Same calibrators
Same sample and reagent volumes
Rate vs. end-point reaction
If the issue affects one assay:
Verify assay parameters with the assay-specific package
insert
Was a different bottle of reagent, calibrator, or control
used, and was it handled correctly?
Was a different lot of calibrator or reagent used?
Are there any non-Abbott assays on the original line? If so,
were carryover studies performed between those assays
and the affected assay?
Move the assay to an alternate line. If the issue is not
resolved, proceed to step 2 of Instrument
Troubleshooting.
1. Is the issue on both lines? If yes, troubleshoot common
items, including:
Lamp
Cuvette washer
Water
Bulk solution use
Maintenance
Probe/mixer wash flow
®
AEROSET
System Troubleshooting Guide
94816-107—November 2004