Unit Does Not Work (No Movement, No Lights): Fault 20.0

SYSTEM FAN RUNNING
6-48
UNIT DOES NOT WORK
(NO MOVEMENT, NO LIGHTS)
CHECK THAT CONTROL
MODULE IS PRESENT &
PLUGGED IN
CHECK THAT ALL SYSTEM
CONNECTIONS ARE PLUGGED IN
CHECK FUSE
CHECK FOR 12/14 VOLTS
ACROSS C1 ON POWER
SUPPLY BOARD
REPLACE POWER
SUPPLY BOARD
REPLACE COMPUTER
BOARD
REPLACE TEST
CONTROL MODULE
Figure 6.20: Unit Does Not Work (No Movement, No Lights)
SYSTEM FAN NOT RUNNING
CHECK THAT UNIT IS
CHECK THAT POWER
POWER SWITCH IS ON
CHECK THAT J3 & J7
ARE PLUGGED IN ON
POWER SUPPLY BOARD
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Section 6
PLUGGED IN
SOURCE IS ACTIVE
CHECK THAT UNIT
REPLACE POWER
SUPPLY BOARD