Page 3
Addendum (06N19-01) and the VP 2000 Processor Operations Manual. This Addendum is issued by Abbott Molecular to provide you with the most current VP 2000 Processor information. Use it in conjunction with VP 2000 Processor software version 1.1 or higher, and the VP 2000 Processor Operations Manual, List Number 03N27-31.
Page 5
Instrument Labeling Symbols Used on Instrument Labeling The symbols in the following table are used on the instrument’s labeling. Table 0.1: Symbols Used on Instrument Labeling Symbol Definition Symbol Definition List number/ Caution or Catalog number Warning Serial number Caution or Warning: Electrical shock Batch code/...
Page 6
Symbol Definition Symbol Definition Manufacturer Indicates the ON Abbott position of a Molecular push-push main Des Plaines, IL power switch 60018 USA Indicates the Warning: OFF position of Pinchpoint a push-push main power switch ABBOTT Waste Electrical MAX-PLANCK- and Electronic...
Page 7
The batteries in this product should be removed at the end of the life of the device by an Abbott Service Technician or a qualified individual, and disposed in accordance with local regulations for separate collection of spend batteries.
Page 8
Safety and Electromagnetic Compatibility (EMC) Standards This section addresses the safety and EMC compliance of the following: • the Abbott VP 2000 Processor • the System Control Center Abbott VP 2000 Processor U.S. and Canadian Safety Standards This instrument has been tested to and complies with standard...
Page 9
European Safety and EMC Standards Safety This instrument meets European requirements for safety (Low Voltage Directive 73/23/EEC). This instrument has been tested to and complies with standards IEC 61010-1:2001, “Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General Requirements” and IEC 61010-2-101, “Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-101: Particular Requirements for In-Vitro Diagnostic...
LED will still be Steady ON even though the basin heater is actually paused, waiting for the basin to cool down to the programmed temperature range. Abbott VP 2000™ Processor Operations Manual Addendum Version 2 30-602111/R1—June 2010...
Page 14
OFF). If this occurs, the basin heater will be Slow turned off automatically. If slow flashing is flashing observed, the VP2000 processor should be powered down and a Field Service Representative should be contacted for assistance. Abbott VP 2000™ Processor Operations Manual Addnedum Version 2 30-602111/R1—June 2010...
Use a plastic cover to protect the instrument against dust and humidity • Store the instrument in its original packaging • Do not stack other materials or goods on the Abbott VP 2000 Processor cardboard packaging • Keep and store the following documents with the instrument: –...
Page 16
Operational Precautions and Limitations Section 7 Operational Precautions and Limitations NOTES Abbott VP 2000™ Processor Operations Manual Addendum Version 2 30-602111/R1—June 2010...
Identifies an activity or area where you may be exposed to hot surfaces. For more information, see Hot Objects, in Physical Hazards in Section 8, Hazards, of the VP 2000 Processor Operations Manual. Abbott VP 2000™ Processor Operations Manual Addendum Version 2 30-602111/R1—June 2010...
Page 18
VP 2000 Processor Operations Manual. CAUTION: Electrostatic Discharge Hazard Observe precautions for handling electrostatic discharge sensitive devices. Wear a wrist ground strap. Discharge static electricity from your body before touching instrument. Abbott VP 2000™ Processor Operations Manual Addendum Version 2 30-602111/R1—June 2010...