Maintenance and Diagnostic Procedures
5-12
1. Open the Charcoal Filter canister by unlatching the top of
the canister.
2. Carefully remove the old Charcoal Cassette and seal in a
bag. Be careful not to spill loose charcoal particles.
3. Dispose of the old Charcoal Cassette according to the
accepted method at the institution.
4. Carefully remove the replacement Charcoal Cassette from
its packaging and set it in the Charcoal Filter canister.
5. Replace the top of the Charcoal Filter canister and close
the latches.
Figure 5.1: Charcoal Cassette
Section 5
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008