Sensor Bar Cleaning Procedure

Required Tools and Consumables
Procedure
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Scouring pad (no soap)
Paper towels
70% Isopropanol alcohol
1. Open the Heated Module Cover.
2. Spray 70% Isopropanol alcohol onto a paper towel and
wipe off the affected probe(s).
3. Lightly wipe the scouring pad over the corroded area.
Figure 5.21: Using Scour Pad to Clean Sensor Bar
4. Wipe the probe to remove any residue with a clean paper
towel.
5. Close the Heated Module Cover.
Sensor Bar Cleaning Procedure
5-33