Repair, Replacement of Spare Parts

Installation

7-70
In order to properly adjust the Interlock Switch, reassembly
must be done slightly different than disassembly.
1. Reverse the operations in steps 5, 4, 3 above. Make sure the
switch is oriented properly, the plunger should be off
center and toward the bottom of the bracket. Install the
spade lugs on the switch as shown in Figure 7.37.
2. Install the hoods.
3. Install the Interlock Switch and adjust it so it is activated
when the Hood is closed and not activated when the Hood
is raised. The switch is adjusted by moving it in or out,
along the slotted mounting bracket. Test the position of
the switch by raising and lowering the Hood. There is an
audible click when the switch is activated.
4. Install the hoods.
CAUTION: Before sliding the Top Cover back into
position, make sure all wiring, cables and connectors
are replaced snugly, within the unit and not hanging
out where they might be sheared off when replacing
the cover.