Abbott VP 2000 Service Manual page 179

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Section 6
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
FLUID WILL NOT RECIRCULATE
KEYPAD RECIRCULATE INDICATOR
LIGHT ON
YES
NO
ROBOTIC ARM IN BATH 2 & DOWN
YES
NO
CHECK CONNECTORS
PASS
FAIL
Figure 6.13: Fluid Will Not Recirculate
Maintenance Diagnostic Faults
ACTIVATE FLOW BUTTON
(RED LIGHT WILL GO ON)
YES
REPLACE CONTROL
MOVE TO BATH 2 & DOWN
REALIGN & PLUG IN
POWER SUPPLY BOARD
TEST BY SUBSTITUTION
COMPUTER BOARD
TEST BY SUBSTITUTION
REPLACE PUMP ASSY
NO
MODULE
6-37