Slide Basket Not Aligned With Vessels: Fault 12.0 - Abbott VP 2000 Service Manual

Processor
Hide thumbs Also See for VP 2000:
Table of Contents

Advertisement

Troubleshooting

Slide Basket Not Aligned With Vessels: Fault 12.0

6-34
The Robotic Arm has two (2) #6-32 socket head set screws,
one in the front and one in the rear, located directly
below the front and rear rotation sensors.
Adjusting the set screws clockwise or counter-clockwise
will shift the stop position of the arm for optimum
alignment with the bath vessels. Apply locking agent to
threads after adjustment is made.
SLIDE BASKET NOT ALIGNED WITH VESSELS
INSPECT BASKET FOR
DEFORMATION
PASS
FAIL
INSPECT ROBOTIC ARM FOR
DEFORMATION
PASS
FAIL
ROTATION ROBOTIC ARM ONLY
ADJUST ROTATION STOPS
Figure 6.11: Slide Basket Not Aligned with Vessels
REPLACE SLIDE BASKET
REBEND INTO
CORRECT SHAPE
1
Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Section 6