To Perform Quasi-Sampling Measurement; To Interrupt Command Execution - Keysight Technologies E5260 Series Programming Manual

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Table 1-5
Programming Basics
Programming Tips

To Perform Quasi-Sampling Measurement

The following setup enables to perform a quasi-sampling measurement. Then the
sampling interval will be sum of delay time and step delay time.
Sets the sweep measurement mode (MM 2 or MM 16).
Sweep start value = Sweep stop value (for WI, WV, or WNX).
Sets hold time, delay time, and step delay time (WT).
Voltage or current
Start value, Stop value
Previous value

To Interrupt Command Execution

The E5260/E5270 executes commands in the received order. However, only the
following commands can interrupt the command execution.
Interrupt Commands
Command
AV
Changes the number of averaging samples during the
measurement.
AIT
Changes the integration time during the measurement.
AB
Aborts the command execution.
*RST
Resets the E5260/E5270 during the command execution.
XE
If the E5260/E5270 has been set to the wait status by the PA or
PAX command, the XE command can be used to release the
wait status. For details, refer to
Reference."
1- 46
Keysight E5260/E5270 Programming Guide, Edition 4
Delay
Hold
Sampling interval = Delay + Step delay
Measurement trigger (e.g. XE)
Description
Chapter 4, "Command
Step delay
: Measurement