Execution
Conditions
Syntax
Parameters
Remarks
Example
Statements
TSC
The TSC command enables or disables the time stamp function.
This command is effective for ASCII data output format. Refer to
4-59.
Time stamp function is not available for the following measurement modes:
Quasi-pulsed spot measurement (MM 9)
Linear search measurement (MM 14)
Binary search measurement (MM 15)
TSC mode
mode :
Time stamp function mode. Integer expression.
mode
0
1
When the function is enabled, the E5260/E5270 returns the time data
with the measurement data. The time data is the time from when the
timer is cleared until the measurement is started. Refer to
Format" on page
To read the time data with the best resolution (100 μs), the timer must be cleared
every 100 sec or less for the FMT1, 2, or 5 data output format, or
every 1000 sec or less for the FMT 11, 12, 15, 21, 22, or 25 data output format.
OUTPUT @E5270;"TSC 1"
Keysight E5260/E5270 Programming Guide, Edition 4
Description
Disables the time stamp function. Initial setting.
Enables the time stamp function.
1-22.
Command Reference
Command Reference
"FMT" on page
"Data Output
4- 121