Keysight Technologies E5260 Series Programming Manual page 334

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

NOTE
Execution
Conditions
Syntax
Example
Statement
Command Reference
Command Reference
XE
The XE command triggers the E5260/E5270 to start measurement, or causes the
E5260/E5270 to recover from the wait state set by the PA command.
This command is not available for the high-speed spot measurement.
After measurement, the measurement data will be entered to the output data buffer.
For data output format, refer to
The following execution conditions are for you who use the XE command to start
measurement. There is no execution condition when you use the XE command to
recover from the wait state.
If any channel is set to the HIGH VOLTAGE state (forcing more than ±42 V, or
voltage compliance set to more than ±42 V) after the trigger (XE), the interlock
terminal must be shorted.
The following commands must be entered before the XE command.
Measurement Mode
Spot
Staircase sweep
Pulsed spot
Pulsed sweep
Staircase sweep with pulsed bias
Quasi-pulsed spot
Liner search
Binary search
Multi channel sweep
XE
OUTPUT @E5270;"XE"
4- 150
Keysight E5260/E5270 Programming Guide, Edition 4
"Data Output Format" on page
CN, MM, DV or DI
CN, MM, WV or WI
CN, MM, PV or PI
CN, MM, PWV or PWI
CN, MM, WV or WI, PV or PI
CN, MM, BDV
CN, MM, LSV or LSI, LGV or LGI
CN, MM, BSV or BSI, BGV or BGI
CN, MM, WI or WV, WNX
1-22.