IN
The IN command sets the specified channel to 0 V with an output range change.
IN [chnum[,chnum...[,chnum]...]]
A maximum of eight channels can be set.
chnum :
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
If you specify multiple chnums, the E5260/E5270 sets the channels to 0 V in the
specified order.
If you do not specify chnum, this command sets all channels to 0 V in order from
higher voltage range (output or measurement range) to lower voltage range.
The IN command sets the specified channels to the following conditions, which are
the same as the conditions after executing the CN command.
Item
Source Mode
Output Voltage
V Range
I Compliance
I Range
Filter
Power Consumption
OUTPUT @E5270;"IN"
OUTPUT @E5270;"IN 1,2,3,5,6"
4- 62
Keysight E5260/E5270 Programming Guide, Edition 4
Table 4-1 on page
4-9.
SMU
V
0 V
20 V
100 μA
100 μA
Not changed
0 W